Improvement of the Detection of Plasmodium vivax by Using iRBC QFlag Values from the Sysmex Hematology Analyzer XN-1000 Coupled with Platelet Count

Authors

  • Sasis Boonmee Clinical Hematology and Microscopy, Medical Technology Department, Maesot Hospital, Tak Province, Thailand

Keywords:

iRBC QFlag, Sysmex XN – 1000, P. vivax, Platelets

Abstract

Plasmodium vivax malaria is the most prevalent in Thailand, and the prompt and accurate detection of the infection are crucial for controlling the disease pandemic. This study aimed to study the efficiency in the diagnosis of P. vivax malarial infection by using the “iRBC QFlag”, a novel parameter developed and obtained from the Sysmex XN-1000 automated analyzer coupled with the platelet count. This is an experimental comparative research was conducted between August 1st, 2022 and October 31st, 2023. From the total of 3,818 samples, there were 1,279 cases with request for CBC and thick film tests, while 2,539 cases with request for CBC, but not thick film test. The iRBC QFlag (≥ 10 AU) and platelet count (< 140 x 109/L) were used to detect P. vivax in blood samples, and the area under the ROC curve was at 0.991 and 0.981, respectively. When using combination of iRBCQFlag (≥10 AU) and platelet count, the sensitivity, specificity, and positive and negative predictive values were 97, 100, 100 and 99.8, respectively. In conclusion, the sensitivity of P. vivax detection increased to 97% when the iRBC QFlag value for positive cutoff was decreased and applied with the platelet counts. This finding suggest that when the value of iRBC QFlag ≥ 10 AU and PLT < 140 x 109/L or iRBC QFlag value of 20-300 AU and PLT ≥ 140 x 109/L, there may be the infection of P. vivax malaria, which should be confirmed by the thick blood film method.

References

Jetsadawut B. Malaria situation and patients’ characteristics in Thasongyang District, Tak Province, Fiscal Years 2018-2022. J of Pub Health Tak Province 2023; 3. (in Thai)

Center of Vector Borne Disease Control 2,3 (Mae Sot), Tak Province. Malaria patient accumulative in Tak Province. 2023. Available from: https://www. facebook.com/photo/?fbid=66198077937 2328&set=a.422537509983324

Shaikh OH, Ahmad SM, Abbasi A, Malik SA, Sahito AA, Munir SM. Thrombocy- topenia in malaria. J Coll Physicians Surg Pak 2009; 19: 708-10.

Marcus VGL, Maria PGM, Helena CCC, Joao BS. Thrombocytopenia in malaria: who cares? Mem Inst Oswaldo Cruz, Rio de Janeiro 2011; 106: 52-63.

Angesom G, Yonas E, Daniel GF, Gebru H, Temesgen F. Thrombocytopenia as a diagnostic marker for malaria in patients with acute febrile illness. J Trop Med 2021; 2021: 5585272. doi: 10.1155/2021/ 5585272.

XN-Series “pRBC?” flagging perfor- mance, Automated Hematology Analyzer XN-Series. Sysmex Corportion.

Hee JH, Jae-Woo C, Seong YP, Seok-Lae C. Vivax malaria detection using a para- sitic red blood cell flag generated by the Sysmex XN-9000 hematology analyzer. Int J Lab Hematol 2021; 43: 1403-7.

Division of vector borne diseases. Guidelines for malaria elimination work for medical and public health personnel in Thailand. Printed no. 1. Bangkok: Aksorn graphic and design publishing; 2019. (in Thai)

Rapepun C. Sample size. Thai Bull Pharm Sci 2007; 4: 1-15. (in Thai)

Haldun A. User’s guide to sample size estimation in diagnostic accuracy studies. Turk J Emerg Med 2022; 22: 177-85

Sackett DL, Haynes RB, Guyatt GH, Tugwell P. Clinical epidemiology: A basic science for clinical medicine. 2ed. New York: Little Brown and co; 1991.

Zou KH, O’Malley AJ, Mauri L. Receiver- Operating Characteristic analysis for evaluating diagnostic tests and predictive models. Circulation 2007; 115: 654-7.

Clinical hematology and microscopy. Maesot Hospital. Standard Operating Procedure: CBC examination 2023.

Shanaz K, Kunal S, Monisha S, Dia M. Rapid detection of Plasmodium vivax by the hematology analyzer for population screening. Diagnostics 2023; 13, 3397; https://doi.org/10.3390/diagnos- tics13223397

Cecile D, Pauline TC, Fanelie M, et al. Flagging performance of Sysmex XN-10 hematology analyser for malaria detection. J Clin Pathol 2020; 73: 676-7.

Downloads

Published

2024-08-13

How to Cite

1.
Boonmee S. Improvement of the Detection of Plasmodium vivax by Using iRBC QFlag Values from the Sysmex Hematology Analyzer XN-1000 Coupled with Platelet Count. วารสารเทคนิคการแพทย์ [internet]. 2024 Aug. 13 [cited 2026 Jan. 4];52(2):9083-94. available from: https://he01.tci-thaijo.org/index.php/jmt-amtt/article/view/267773

Issue

Section

Original Articles