Validation of reprocessing process for breathing circuits used in ventilator and high-flow nasal cannula in a hospital

Authors

  • Athicha Kittiwattanokhun National Institute of Metrology (Thailand)
  • Natthakit Thirakittithanawong Phrapokklao Hospital, Chanthaburi Provinc
  • Thanakorn Kwanchalernjai Phrapokklao Hospital, Chanthaburi Provinc
  • Kitiya Fungprayun Phrapokklao Hospital, Chanthaburi Province
  • Anong Anuwatthananon Phrapokklao Hospital, Chanthaburi Province
  • Supattra Pongsak Phrapokklao Hospital, Chanthaburi Province
  • Thanwa Pitaksuteepong Phrapokklao Hospital, Chanthaburi Province
  • Theera Leeudomwong Faculty of Applied Science King Mongkut’s University of Technology North Bangkok
  • Tassanai Sanponpute National Institute of Metrology

Keywords:

reprocessing, breathing circuit, residual protein and blood testing, biological safety, leakage testing

Abstract

Phrapokklao Hospital currently consumes a large number of disposable medical devices, including critical equipment such as ventilator breathing circuits and high-flow nasal cannula (HFNC) circuits. These devices are typically discarded after a single use to minimize the risk of cross-infection, owing to concerns that conventional reprocesses may not completely eliminate microbial contamination. This study aimed to evaluate and validate the safety of reprocessing both reusable and single-use breathing circuits. A sample of swivel elbow connectors and tubes were tested for residual protein and blood after cleaning, followed by sterility and cytotoxicity assessments after sterilization. In addition, leak testing and visual inspections were performed on single use ventilator breathing circuits and HFNC circuits from three different brands currently used in the hospital.

The study revealed no residual protein or blood, no microbial growth, and no cytotoxic effects in the tested samples. Leak testing and visual inspections confirmed that the circuits remained in acceptable working condition, with only a slight increase in leakage observed after ten reuse cycles. These findings indicate that These findings indicate that single-use breathing circuits can be safely reused through effective reprocessing procedures, thereby substantially reduce both medical waste and healthcare costs without compromising patient safety. Nevertheless, further studies are required to validate long-term outcomes and to determine the broader applicability of this approach in clinical practice.

References

Li J, Fink JB, Ehrmann S. High-flow nasal cannula for COVID-19 patients: low risk of bio-aerosol dispersion. Eur Respir J 2020;55(5):2000892.

McGain F, Story D, Lim T, McAlister S. Financial and environmental costs of reusable and single-use anaesthetic equipment. Br J Anaesth 2017;118(6):862-9.

Wijayanti LE, Adisasmito WB. The effectiveness of reprocessing single-use medical devices in improving cost efficiency and hospital resource management: a systematic review. Asian J Eng Soc Health 2025;4(8):1304-12.

Boonchaisri P, Lertwatthanawilat W, Unahalekhaka A. Reuse of single-use medical devices in secondary and tertiary hospitals. Int J Evid Based Healthc 2016;14:p S15.

Rudhart SA, Günther F, Dapper LI, Gehrt F, Stuck BA, Hoch S. Analysis of bacterial contamination and the effectiveness of UV light-based reprocessing of everyday medical devices. PLoS One 2022;17(11):e0268863.

Barmouz M, Azarhoushang B, Kintzel W, Bucher V. Long-term impact of sterilization cycles on the surface and mechanical integrity of medical-grade silicone. J Manuf Mater 2025;9(8):282.

World Health Organization. Decontamination and reprocessing of medical devices for health-care facilities: aide-memoire. Geneva: World Health Organization; 2022. Report No.: WHO/UHL/IHS/IPC/2022.4.

Food and Drug Administration (US). Reprocessing medical devices in health care settings: validation methods and labeling. Guidance for industry and FDA staff. Silver Spring (MD): Food and Drug Administration; 2015. Appendix E updated 2017 Jun.

Gruber S, Nickel A. Toxic or not toxic? The specifications of ISO 10993-5 are not explicit enough to yield comparable results in cytotoxicity assessment of an identical medical device. Front Med Technol 2023;5:1195529.

United States Pharmacopeia. Sterility tests 〈71〉 [Internet]. Rockville (MD): United States Pharmacopeia; 2017 [cited 2024 Aug 17]. Available from: https://doi.usp.org/USPNF/USPNF_M98810_01_01.html

Thai Industrial Standards Institute. TIS 2272-2551: Endotracheal tubes. Bangkok: Thai Industrial Standards Institute, Ministry of Industry; 2017.

International Organization for Standardization. ISO 10993-5: Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity. Geneva: ISO; 2023.

International Organization for Standardization. ISO 5367:2023. Anaesthetic and respiratory equipment-Breathing sets and connectors. Geneva: ISO; 2023.

Tipple AFV, Sodré RLR, Nascimento LC, Costa DM. Reusable medical device precleaning in care units: What are the indicators to prevent biofilm formation and control occupational biological risk? Hygienes 2024;4(1):115-21.

Kremer TA, Patel A, Summers C, Quin M, Lemons K, McDonnell G. Protein residuals on reusable medical devices and patient safety impact. Zentr Steril 2019;27(3):178-83.

Uematsu M, Miyamoto Y, Shimizu M, Kajiura T, Saito A, Takashina, M, et al. Design and validation of a method for evaluating medical device cleanliness by recovering and quantifying residual proteins on stainless plates. Sci Rep 2024;14(1):21982.

Huang J, Yi L, Chen Y, Hu J. Factors associated with cleaning quality of reusable medical devices at a single center in China. Heliyon 2024;10(2):e24194.

Kwon J, Kang E, Shin S, Lee B, Ko M, Kim S, et al. Ventilation difficulty caused by obstructed heated breathing circuit. Medicina 2023;59(5):965.

Downloads

Published

2026-06-28

How to Cite

1.
Kittiwattanokhun A, Thirakittithanawong N, Kwanchalernjai T, Fungprayun K, Anuwatthananon A, Pongsak S, Pitaksuteepong T, Leeudomwong T, Sanponpute T. Validation of reprocessing process for breathing circuits used in ventilator and high-flow nasal cannula in a hospital. J Nurs Sci Health [internet]. 2026 Jun. 28 [cited 2026 Sep. 4];49(2):205-20. available from: https://he01.tci-thaijo.org/index.php/nah/article/view/282158

Issue

Section

Research Article