Agreement of Intraocular Pressure Measurements between Ophthalmologist and Ophthalmic Nurse Using Goldmann Applanation and Icare Rebound Tonometry Methods in Eye Clinic of Phetchabun Hospital
Keywords:
icare tonometer, goldmann applanation tonometry (gat), intraocular pressure (iop), corneal thicknessAbstract
Objective: This study compared the consistency of intraocular pressure measurements between the iCare rebound tonometer, measured by ophthalmic nurses, and the standard Goldmann Applanation Tonometry (GAT), measured by ophthalmologists. Methods: Prospective Cross-sectional hospital based study. Results: This study enrolled 225 eyes from 113 patients to compare intraocular pressure (IOP) measurements obtained using rebound tonometry (iCare IC200) and Goldmann applanation tonometry (GAT). Intraclass correlation coefficients (ICCs) were calculated to assess inter-method reliability. Lin's concordance correlation coefficient, along with Bland-Altman analysis, was employed to evaluate the relationship and agreement between the two measurement methods. The findings indicated excellent agreement between iCare tonometry and GAT. Furthermore, corneal thickness did not significantly influence the consistency of IOP measurements for either method. Conclusion: This study indicates that the iCare tonometer is a reliable tool for measuring intraocular pressure, comparable to the standard GAT method, and may be beneficial for glaucoma screening in the community and for measuring intraocular pressure in situations where the GAT method is limited.
References
Brusini P, Salvetat ML, Zeppieri M. How to Measure Intraocular Pressure: An Updated Review of Various Tonometers. J Clin Med 2021;10(17):3860.
Shan S, Wu J, Cao J, Feng Y, Zhou J, Luo Z, et al. Global incidence and risk factors for glaucoma: A systematic review and meta-analysis of prospective studies. J Glob Health 2024 Nov 8:14:04252.
Sihota R, Angmo D, Ramaswamy D, Dada T. Simplifying “target” intraocular pressure for different stages of primary open-angle glaucoma and primary angle-closure glaucoma. Journal of Ophthalmology 2018;66(4):495–505.
Tomita R, Sharpe GP, Betsch D, Bonatti R, Chauhan BC. A Comparative Study of the Handheld IC200 and Slit Lamp-mounted ST500 Rebound Tonometers with Goldmann Applanation Tonometry. Ophthalmol Glaucoma 2025;8(2):152-6.
Badakere SV, Chary R, Choudhari NS, Rao HL, Garudadri C, Senthil S. Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea. Ophthalmol Glaucoma 2021;4(3):238–43.
Sachdeva R, Iordanous Y, Lin T. Comparison of intraocular pressure measured by iCare tonometers and Goldmann applanation tonometer. Can J Ophthalmol 2023;58(5):426–32.
Ozcura F, Yildrim N, Sahin A, Colak E. Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes. Int J Ophthalmol 2015;8(2):299–304.
American Academy of Ophthalmology. Goldmann Tonometer Remains Gold Standard for Measuring IOP. [Internet]. 2017 [cited 2025 Feb 18]. Available from: https://shorturl.asia/A65zL
Krolo I, Mihaljevic B, Kasumovic A, Bagatin F, Ravlic MM, Herman JS. Rebound Tonometry over Soft Contact Lenses. Acta Inform Med 2020;28(3):185–9.
Shen X, Xu Y, Ye Y, Huai S, Wu P, Huang J, et al. Aerosolization ocular surface microorganisms accumulation effect during non-contact tonometer measurements. BMC Ophthalmol 2024;24(1):392.
Guo H, Li W, Huang Y, Li X, Li Z, Zhou H, et al. Increased microbial loading in aerosols produced by non-contact air-puff tonometer and relative suggestions for the prevention of coronavirus disease 2019 (COVID-19). PLoS One 2020;15(10).
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 สมาคมเวชศาสตร์ป้องกันแห่งประเทศไทย

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
บทความที่ลงพิมพ์ในวารสารเวชศาสตร์ป้องกันแห่งประเทศไทย ถือเป็นผลงานวิชาการ งานวิจัย วิเคราะห์ วิจารณ์ เป็นความเห็นส่วนตัวของผู้นิพนธ์ กองบรรณาธิการไม่จำเป็นต้องเห็นด้วยเสมอไปและผู้นิพนธ์จะต้องรับผิดชอบต่อบทความของตนเอง