The role of medical physicists in the installation and clinical implementation of linear accelerators for radiotherapy: clinical practice and risk management considerations

Authors

  • Jaruek Kanphet Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society
  • Sirinya Ruangchan Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society
  • Nuttha Plangpleng Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society
  • Tanawat Tawonwong Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society
  • Taweap Sanghangthum Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society

Keywords:

medical physicist, linear accelerator, commissioning, acceptance test, quality assurance

Abstract

The installation and clinical implementation of a linear accelerator (LINAC) for radiotherapy are complex processes that require close collaboration among multidisciplinary healthcare professionals. Medical physicists play a pivotal role throughout the entire project lifecycle, from infrastructure planning and radiation shielding design to acceptance testing, radiation surveys, beam commissioning, treatment planning system (TPS) modeling, and pre-clinical quality assurance verification. These activities are essential to ensure the accuracy, safety, and reliability of radiation treatment delivery.

This article aims to share practical experience and provide guidance for early-career medical physicists involved in installing and commissioning new LINAC systems. The discussion is based on recommendations and professional guidelines published by the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the International Atomic Energy Agency (IAEA). A clinical governance framework is presented, encompassing four major phases: strategic planning and infrastructure design, installation and acceptance testing, beam commissioning and TPS modeling, and clinical validation before service release.

In addition, this article highlights key aspects of risk management, quality assurance, documentation, and regulatory compliance that contribute to the successful implementation of a new radiotherapy facility. The evolving role of medical physicists as leaders in clinical safety, quality management, and technological integration is also discussed. By sharing practical insights and lessons learned, this article seeks to support the professional development of medical physicists and promote safe, effective, and high-quality radiotherapy services.

References

1. Khan FM, Gibbons JP. Khan’s The Physics of Radiation Therapy. 6th ed. Wolters Kluwer; 2019.

2. McGinley PH. Shielding techniques for radiation oncology facilities. 3rd ed. Madison (WI): Medical Physics Publishing; 2010.

3. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37(8):4078-101.

4. International Atomic Energy Agency. Roles and responsibilities, and education and training requirements for clinically qualified medical physicists. Vienna: IAEA; 2013. (IAEA Human Health Series No. 25).

5. Clements JB, Baird CT, de Boer SF, Fairobent LA, Fisher Tyler, Goodwin JH, et al. AAPM medical physics practice guideline 10.a.: scope of practice for clinical medical physics. J Appl Clin Med Phys. 2018;19:11-25.

6. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al. Task Group 142 report: Quality assurance of medical accelerators. Med Phys. 2009;36:4197–4212.

7. ESTRO. ESTRO Booklet No. 9: Guidelines for the verification of IMRT. Brussels: European Society for Radiotherapy and Oncology; 2008.

8. International Atomic Energy Agency. Radiation oncology physics: a handbook for teachers and students. Vienna: IAEA; 2005.

9. National Council on Radiation Protection and Measurements. Structural shielding design and evaluation for megavoltage X- and gamma-ray radiotherapy facilities. Bethesda (MD): NCRP; 2005. (NCRP Report No. 151).

10. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317.

11. Das IJ, Cheng CW, Watts RJ, Ahnesjo A, Gibbons John, Li XA, et al. Accelerator beam data commissioning equipment and procedures: report of the TG-106 of the Therapy Physics Committee of the AAPM. Med Phys. 2008;35:4186-215.

12. “International Atomic Energy Agency”. Absorbed dose determination in external beam radiotherapy. Vienna: IAEA; 2000. (Technical Reports Series No. 398).

13. Alfonso R, Andreo P, Capote R, Saiful Huq M, Kilby W, Kjall P, et al. A new formalism for reference dosimetry of small and nonstandard fields. Med Phys. 2008;35:5179-86.

14. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656-61.

15. Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36:5359-73.

16. Mynampati DK, Yaparpalvi R, Hong L, Kuo HC, Mah D. Application of AAPM TG 119 to volumetric arc therapy (VMAT). J Appl Clin Med Phys. 2012;13:338-48.

17. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki Todd, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: recommendations of AAPM Task Group 218. Med Phys. 2018;45:e53-e83.

18. Ford EC, Gaudette R, Myers L, Vanderver B, Engineer L, Zellars R, et al. Evaluation of safety in a radiation oncology setting using failure mode and effects analysis. Int J Radiat Oncol Biol Phys. 2009;74:852-8.

19. International Atomic Energy Agency. Lessons learned from accidental exposures in radiotherapy. Vienna: IAEA; 2000. (Safety Reports Series No. 17).

20. Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581-618.

21. Yan D, Vicini F, Wong J, Martinez A. Adaptive radiation therapy. Phys Med Biol. 1997;42:123-32.

22. Hussein M, Heijmen BJM, Verellen D, Nisbet A. Automation in intensity modulated radiotherapy treatment planning—a review of recent innovations. Br J Radiol. 2018;91:20180270.

Downloads

Published

2026-10-28

How to Cite

1.
Kanphet J, Ruangchan S, Plangpleng N, Tawonwong T, Sanghangthum T. The role of medical physicists in the installation and clinical implementation of linear accelerators for radiotherapy: clinical practice and risk management considerations. J Thai Assn of Radiat Oncol [Internet]. 2026 Oct. 28 [cited 2026 Nov. 27];32(2):O1-O12. Available from: https://he01.tci-thaijo.org/index.php/jtaro/article/view/286316

Issue

Section

Review articles