Accuracy of radiation dose for cancer treatment
Abstract
Objective: The accuracy of radiation dose will reduce the risk of radiation complication to the patients. The accuracy of radiation dose at Songklanagarind hospital was audited by the postal TI.D intercomparision which is the IAEA standard method.
Materials and method: The radiation output of Cobalt-60 machine measured and calculated the irradiation time for 200 cGy in water to TLD by the LAEA setup conditions. Expose 200 cGy to each TLDs by the calculated irradiation time. These exposed TLDs were proceed at the SSDL of Medical Science Department and reported to the hospital. Evaluation data were collected from 1984-2002.
Discussion of results: The radiation dose of exposed TLDs will be interpreted by Medical Science Department using IAEA standard scoring. The percentage difference from the quoted radiation of more than 5% would be requested more data, repeat the measurement or need to visit the site by the expert physicists.
Conclusion: The TLD evaluation were 11 satisfied, 2 were acceptable and 2 were request more data. The maximum degree of different was (-5.45%) and minimum was (-0.05%). The treatment time calculation software was developed to reduce the degree of calculation error. Result of TLD evaluation was used to check the dose accuracy agreement of the ionization chamber measurement and treatment planning system calculation. The results could support the physicist accurately control the radiation dose in patient treatment and the understanding of radiation dosage. The site could know self accuracy and have a qualified reference of radiation dose. The radiation risk reduction for patient treatment is extremely required for the evaluation purpose.
References
วิชาญ หล่อวิทยา, Principles of Radiation Oncology, In: วิชาญ หล่อวิทยา, ไพรัช เทพมงคล, ประมุข พรหม รัตนพงศ์, ชนะวัธน์ เทศะวิบุล, Manual of Radiation Oncology. โรงพิมพ์แห่งจุพาลงกรณ์มหาวิทยาลัย, 2544,12.
McKinzie A, Level of detail required in physics quality system, In: Mayles R. Physics of Quality Control in Radiotherapy, The Institute of Physics and Engineering in Medicine, England, 19999, 28.
Izewska J, Andreo P, The IAEA/WHO TLD postal programme for radiotherapy Hospital, Radiother Oncol, 2000; 54: 65-72.
Izewska J, Svensson H, Ibbott G, Worldwide QA networks for radiottherapy dosimetry. IAEA-CN-96-76.
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
บทความที่ได้รับการตีพิมพ์เป็นลิขสิทธิ์ของวารสารมะเร็งวิวัฒน์ ข้อความที่ปรากฏในบทความแต่ละเรื่องในวารสารวิชาการเล่มนี้เป็นความคิดเห็นส่วนตัวของผู้เขียนแต่ละท่านไม่เกี่ยวข้องกับ และบุคคลากรท่านอื่น ๆ ใน สมาคมฯ แต่อย่างใด ความรับผิดชอบองค์ประกอบทั้งหมดของบทความแต่ละเรื่องเป็นของผู้เขียนแต่ละท่าน หากมีความผิดพลาดใดๆ ผู้เขียนแต่ละท่านจะรับผิดชอบบทความของตนเองแต่ผู้เดียว