Performance evaluation of myQA Dailyfor output constancy check on linear accelerator
Keywords:
Linear accelerator, myQA Daily, Quality assuranceAbstract
Background: The myQA Daily device is a daily quality control device for controlling machine-specific parameters, such as dose output, flatness, symmetry, center, field size, and energy. The myQA Daily device also contains temperature sensors and pressure sensors for measuring temperature and pressure.
Objective: To evaluate the performance of the myQA Daily device as a daily quality control device for a linear accelerator.
Materials and Methods: Performance evaluation was performed using Varian TrueBEAM linear accelerator at 6 MV photon beams and 9 MeV electron beams. The evaluating parameters consist of short-term reproducibility, linearity, temperature and pressure, dose rate and output constancy. Measurements of the machine using the myQA daily device were compared to ionization chamber, thermometer and barometer.
Results: The coefficient of variation for the myQA Daily short-term reproducibility was 0.05% and also showed a linear dose output response. On temperature and pressure test, the myQA Daily agreed with calibrated thermometer and barometer. The variation of dose rate was within 0.30%. Over the period, the myQA Daily output constancy agreed with the ionization chamber to within 1.30%.
Conclusion: The myQA Daily device is reliable as an ionization chamber for routine quality assurance of linear accelerator.
References
สาขารังสีรักษา และมะเร็งวิทยาคณะแพทยศาสตร์ จุฬาลงกรณ์มหาวิทยาลัย. รังสีรักษา. 2559 [สืบค้นเมื่อ 19 สิงหาคม 2563]. จาก https://www.chulacancer.net/.
Klein EE, Hanley J, Bayouth J, Yin F-F, Simon W, Dresser S, et al. Task Group 142 reports: quality assurance of medical accelerator. Med Phys. 2009; 36: 4197-212.
IBA Dosimetry GmbH., Schwarzenbruch, Germany. myQA Daily User's Guide. 2019 [Retrieved July 31, 2020]. Available from: https://www.iba-dosimetry.com/product/myqa-daily/.
Podgorsak E.B. Radiation oncology physics: a handbook for teachers and student. Vienna: International Atomic Energy Agency; 2005.
Michael P Barnes, Peter B Greer. Evaluation of the TrueBeam machine Performance check (MPC) beam constancy checks for flattened and flattening filter-free (FFF) photon beams. J Appl clin Med Phys. 2017; 18: 139-150.
Binny D, Lancaster C.M, Karin T, Trapp J.V, Crowe S.B. Monitoring Daily QA3 constancy for routine quality assurance on linear accelerators. Phys Med. 2016; 32: 1479-1487.
Downloads
Published
How to Cite
Issue
Section
License
บทความที่ได้รับการตีพิมพ์เป็นลิขสิทธิ์ของวารสารมะเร็งวิวัฒน์
ข้อความที่ปรากฏในบทความแต่ละเรื่องในวารสารวิชาการเล่มนี้เป็นความคิดเห็นส่วนตัวของผู้เขียนแต่ละท่านไม่เกี่ยวข้องกับ และบุคคลากรท่านอื่น ๆ ใน สมาคมฯ แต่อย่างใด ความรับผิดชอบองค์ประกอบทั้งหมดของบทความแต่ละเรื่องเป็นของผู้เขียนแต่ละท่าน หากมีความผิดพลาดใดๆ ผู้เขียนแต่ละท่านจะรับผิดชอบบทความของตนเองแต่ผู้เดียว