A Consistency Study of Blood Ferritin Levels between the Preanalytical Dilution Versus Non-Dilution

Authors

  • Worrawan Wasusophaphon Department of Medical Technology and Clinical Pathology, Nakornping Hospital
  • Krittapak Hawjarearn Department of Medical Technology and Clinical Pathology, Nakornping Hospital

Keywords:

Blood ferritin, Preanalytical dilution

Abstract

Background: Analytical for the determination of blood ferritin levels in our laboratory, we had to dilute and re-analyze samples when the value was over the measurement interval. The objective of this study was to study the consistency of blood ferritin levels between the preanalytical dilution versus non-dilution technique.

Method: Cross-sectional diagnostic study was performed in the analysis of the Immunoturbidimetric principle of blood ferritin. There were 2x, 3x, and 4x dilution of the preanalytical dilution compared to the non-diluted group. Data were analyzed with descriptive statistics, Spearman correlation statistics and linear regression analysis for correlations, Paired t-test statistics to test the difference in mean ferritin levels, Wilcoxon signed rank test, Kappa statistics and ROC to test the consistency and assess the accuracy of the interpretations.

Results: A total of 138 samples, there were 84 female (60.9%), 54 males (39.13%), average ages 61-81 years old. The study had found a correlation between blood ferritin levels in the preanalytical diluted and non-diluted samples before analysis. The correlation coefficients (r) in preanalytical diluted samples at 2x, 3x and 4x dilution were 0.994, 0.989, 0.977 (P<0.01), respectively. The difference of the sample dilution at 2x had the least difference (Mean Diff -24.74, P<0.01). When tested with Kappa, ferritin levels were statistically consistent (Kappa=0.898, 95%CI=0.825 – 0.971 (p<0.01). Sensitivity 90.90%, Specificity 100.00%, PPV 100%, NPV 89.7% and AUC= 94.90% (95%CI 0.912 - 0.985). A preanalytical dilution of 2x sample can reduce of diluted and repeated 744 of the 4,012 samples (18.5%). The pre-analysis of 3x and 4x sample dilution, the measurement interval can report value up to 1,500 ng/ml. and 2,000 ng/ml.

Conclusion: A preanalytical dilution of 2x sample was able to report with excellent agreement consistency level. However, there are increased in difference with more dilution samples. Technicians should coordinate with physicians and discuss with clinical scenario.

References

McLean E, Cogswell M, Egli I, Wojdyla D, de Benoist B. Worldwide prevalence of anaemia, WHO Vitamin and Mineral Nutrition Information System, 1993-2005. Public Health Nutr. 2009;12(4):444-54. doi: 10.1017/S1368980008002401.

Department of Mental Health. Iron Deficiency Anemia Resulting in decreasing 5-10 points of IQ in Scholl age [Internet]; c2019. Available from: https://www.dmh.go.th/news-dmh/view.asp?id=29701.

Nakorn SN. Iron Deficiency Anemia. J Hematol Transf Med. 1991;1(1):77-104.

Kidchalard J, Tayapiwatana C. Detection of Serum Ferritin by Indirect Sandwich ELISA. Bull Chiang Mai Assoc Med Sci. 2003;36(2):91-9.

WHO GUIDELINE DEVELOPMENT GROUP. WHO guideline on use of ferritin concentrations to assess iron status in individuals and populations. Geneva: World Health Organization; 2020.

Goodhand JR, Kamperidis N, Rao A, Laskaratos F, McDermott A, Wahed M, et al. Prevalence and management of anemia in children, adolescents, and adults with inflammatory bowel disease. Inflamm Bowel Dis. 2012;18(3):513-9. doi: 10.1002/ibd.21740.

Short MW, Domagalski JE. Iron deficiency anemia: evaluation and management. Am Fam Physician. 2013;87(2):98-104.

Viprakasit V. Comprehensive Management for Thalassemia. Hematol Transfus Med. 2013;23(4):303-20. Available from: http://www.tsh.or.th/file_upload/files/09(4).pdf

Abbott Laboratories. Quantia ferritin package insert. Abbott: Wiesbaden Germany; 2017.

Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et. al. Desirable Specifications for Total Error, Imprecision, and Bias, derived from intra- and inter-individual biologic variation [Internet]; c2014. Available from: https://www.westgard.com/biodatabase1.htm

Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.

Ray P, Manach YL, Riou B, Houle TT. Statistical Evaluation of a Biomarker. Anesthesiology. 2010; 112(4):1023-40. doi: 10.1097/ALN.0b013e3181d47604.

Wu SJ, Hayden JA. Upfront dilution of ferritin samples to reduce hook effect, improve turnaround time and reduce costs. Biochem Med (Zagreb). 2018;28(1):010903. doi: 10.11613/BM.2018.010903.

Dahman LSB, Sumaily KM, Sabi EM, Hassan MA, Thalab AMB, Sayad AS, et al. A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods. Diagnostics (Basel). 2022;12(2):320. doi: 10.3390/diagnostics12020320.

Sarakarn P, Munpolsri P. OPTIMAL CUT-OFF POINTS FOR RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE ANALYSIS IN DEVELOPING TOOLS OF HEALTH INNOVATIONS: EXAMPLE USING STATA. Thai Bull Pharm Sci. 2021;16(1):93-108.