Comparisons of Miniscrew Implant Primary Stability between Two Different Synthetic Bone Densities during Placement with Either Self-Drilling or Self-Tapping Technique: An in vitro Study
Main Article Content
Abstract
The purpose of this study was to investigate the influence of cancellous bone densities on miniscrew implant primary stability during placement with either self-drilling or self-tapping technique. Forty titanium alloy miniscrew implants were divided into two groups, according to their placement techniques, self-drilling and self-tapping, and placed into synthetic bone of different cancellous bone densities: either 0.64 g/cm3 or 0.32 g/cm3. Maximal insertion torque and vertical pull-out strength were recorded in Ncm and N, respectively. Two-way ANOVA was used to detect interaction between miniscrew implant placement technique and cancellous bone density factors. The result showed that there was a significant interaction between miniscrew implant placement techniques and different cancellous bone densities (p<0.001) on the miniscrew implant primary stability parameters. The insertion torque and pull-out strength values were significantly greater for the miniscrew implant placed in the high-density cancellous bone than in the low-density one (p<0.001). Miniscrew implant placement into the same cancellous bone density with self-drilling technique showed significantly greater insertion torque (p<0.001) and pull-out strength (p<0.05) than those with self-tapping technique. However, no statistically significant difference in maximal insertion torque was found between self-tapping miniscrew implants placed in high-density, and self-drilling miniscrew implants placed in low-density. In conclusion, cancellous bone density influences miniscrew implant primary stability for both self-tapping and self-drilling placement systems. Pre-drilling is recommended for miniscrew implant placement into high-density cancellous bone to avoid high insertion torque and maintain adequate pull-out strength.
Article Details
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
บทความ ข้อมูล เนื้อหา รูปภาพ ฯลฯ ที่ได้รับการลงตีพิมพ์ในวิทยาสารทันตแพทยศาสตร์ มหาวิทยาลัยขอนแก่นถือเป็นลิขสิทธิ์เฉพาะของคณะทันตแพทยศาสตร์ มหาวิทยาลัยขอนแก่น หากบุคคลหรือหน่วยงานใดต้องการนำทั้งหมดหรือส่วนหนึ่งส่วนใดไปเผยแพร่ต่อหรือเพื่อกระทำการใด ๆ จะต้องได้รับอนุญาตเป็นลายลักษณ์อักษร จากคณะทันตแพทยศาสตร์ มหาวิทยาลัยขอนแก่นก่อนเท่านั้น
References
Kuroda S, Sugawara Y, Deguchi T, Kyung HM, Takano Yamamoto T. Clinical use of miniscrew implants as orthodontic anchorage: success rates and postoperative discomfort. Am J Orthod Dentofacial Orthop 2007;131(1):9-15.
Wiechmann D, Meyer U, Büchter A. Success rate of mini‐and micro‐implants used for orthodontic anchorage: a prospective clinical study. Clin Oral Implants Res 2007;18(2):263-7.
Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR. Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthop Trauma 2001;15(3):216-21.
Eraslan O, İnan Ö. The effect of thread design on stress distribution in a solid screw implant: a 3D finite element analysis. Clin Oral Investig 2010;14(4):411-6.
Öktenoǧlu BT, Ferrara LA, Andalkar N, Özer AF, Sarioǧlu AÇ, Benzel EC. Effects of hole preparation on screw pullout resistance and insertional torque:a biomechanical study. J Neurosurg Spine 2001;94(1):91-6.
Kim YK, Kim YJ, Yun PY, Kim JW. Effects of the taper shape, dual-thread, and length on the mechanical properties of mini-implants. Angle Orthod 2009;79(5): 908-14.
Gracco A, Giagnorio C, Incerti Parenti S, Alessandri Bonetti G, Siciliani G. Effects of thread shape on the pullout strength of miniscrews. Am J Orthod Dentofacial Orthop. 2012;142(2):186-90.
Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006;67(3):162-74.
Watanabe H, Deguchi T, Hasegawa M, Ito M, Kim S, Takano‐Yamamoto T. Orthodontic miniscrew failure rate and root proximity, insertion angle, bone contact length, and bone density. Orthod Craniofac Res 2013; 16(1):44-55.
Iijima M, Takano M, Yasuda Y, Muguruma T, Nakagaki S, Sakakura Y, et al. Effect of the quantity and quality of cortical bone on the failure force of a miniscrew implant. Eur J Orthod 2012;35(5):583-9.
Motoyoshi M, Yoshida T, Ono A, Shimizu N. Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants 2007;22(5):779-84.
Marquezan M, Mattos CT, Sant'Anna EF, de Souza MMG, Maia LC. Does cortical thickness influence the primary stability of miniscrews?: A systematic review and meta-analysis. Angle Orthod 2014;84(6):1093-103.
Chen Y, Kyung HM, Gao L, Yu W-J, Bae E-J, Kim S-M. Mechanical properties of self-drilling orthodontic micro-implants with different diameters. The Angle Orthodontist 2010;80(5):821-7.
Marquezan M, Lima I, Lopes RT, Sant'Anna EF, de Souza MMG. Is trabecular bone related to primary stability of miniscrews? Angle Orthod 2014;84(3): 500-7.
Çehreli S, Arman-Özçırpıcı A. Primary stability and histomorphometric bone-implant contact of self-drilling and self-tapping orthodontic microimplants. Am J Orthod Dentofacial Orthop. 2012;141(2):187-95.
Hung E, Oliver D, Kim KB, Kyung HM, Buschang PH. Effects of pilot hole size and bone density on miniscrew implants' stability. Clin Implant Dent Relat Res 2012; 14(3):454-60.
Heidemann W, Gerlach KL, Gröbel K-H, Köllner H-G. Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 1998;26(1):50-5.
Nucera R, Lo Giudice A, Bellocchio AM, Spinuzza P, Caprioglio A, Perillo L, et al. Bone and cortical bone thickness of mandibular buccal shelf for mini-screw insertion in adults. The Angle Orthodontist 2017;87(5): 745-51.
Devlin H, Horner K, Ledgerton D. A comparison of maxillary and mandibular bone mineral densities. J Prosthet Dent 1998;79(3):323-7.
Uemura M, Motoyoshi M, Yano S, Sakaguchi M, Igarashi Y, Shimizu N. Orthodontic mini-implant stability and the ratio of pilot hole implant diameter. Eur J Orthod 2012;34(1):52-6.
Ciarelli M, Goldstein S, Kuhn J, Cody D, Brown M. Evaluation of orthogonal mechanical properties and density of human trabecular bone from the major metaphyseal regions with materials testing and computed tomography. J Orthop Res 1991;9(5):674-82.
Cha JY, Kil JK, Yoon TM, Hwang CJ. Miniscrew stability evaluated with computerized tomography scanning. Am J Orthod Dentofacial Orthop 2010; 137(1):73-9.
Marquezan M, Souza MMGd, Araújo MTdS, Nojima LI, Nojima MdCG. Is miniscrew primary stability influenced by bone density? Braz Oral Res 2011; 25(5):427-32.
Pan CY, Liu PH, Tseng YC, Chou ST, Wu CY, Chang HP. Effects of cortical bone thickness and trabecular bone density on primary stability of orthodontic mini-implants. J Dent Sci 2019.
Carano A, Lonardo P, Velo S, Incorvati C. Mechanical properties of three different commercially available miniscrews for skeletal anchorage. Prog Orthod 2005; 6(1):82-97.
Friberg B, Sennerby L, Roos J, Lekholm U. Identification of bone quality in conjunction with insertion of titanium implants. A pilot study in jaw autopsy specimens. Clin Oral Implants Res 1995;6(4): 213-9.
Kim JW, Ahn SJ, Chang YI. Histomorphometric and mechanical analyses of the drill-free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2005;128(2):190-4.
Tepedino M, Masedu F, Chimenti C. Comparative evaluation of insertion torque and mechanical stability for self-tapping and self-drilling orthodontic miniscrews–an in vitro study. Head Face Med 2017; 13(1):10.
Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res 2006;17(1):109-14.
Suzuki EY, Suzuki B. Placement and removal torque values of orthodontic miniscrew implants. Am J Orthod Dentofacial Orthop 2011;139(5):669-78.
Suzuki M, Deguchi T, Watanabe H, Seiryu M, Iikubo M, Sasano T, et al. Evaluation of optimal length and insertion torque for miniscrews. Am J Orthod Dentofacial Orthop 2013;144(2):251-9.
Di Leonardo B, Ludwig B, Lisson JA, Contardo L, Mura R, Hourfar J. Insertion torque values and success rates for paramedian insertion of orthodontic mini-implants. J Orofac Orthop 2018;79(2):109-15.
Gantous A, Phillips JH. The effects of varying pilot hole size on the holding power of miniscrews and microscrews. Plast Reconstr Surg 1995;95(7):1165-9.
Battula S, Schoenfeld A, Vrabec G, Njus GO. Experimental evaluation of the holding power/stiffness of the self-tapping bone screws in normal and osteoporotic bone material. Clinical Biomechanics 2006;21(5):533-7.
Wang Z, Zhao Z, Xue J, Song J, Deng F, Yang P. Pullout strength of miniscrews placed in anterior mandibles of adult and adolescent dogs: a microcomputed tomographic analysis. Am J Orthod Dentofacial Orthop 2010;137(1):100-7.
Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull-out strength of monocortical screws placed in the maxillae and mandibles of dogs.Am J Orthod Dentofacial Orthop 2005;127(3):307-13.
Steeves M, Stone C, Mogaard J, Byrne S. How pilot-hole size affects bone-screw pullout strength in human cadaveric cancellous bone. Can J Surg. 2005;48(3):207.
Defino HL, Wichr CRG, Shimano AC, Kandziora F. The influence of pilot hole diameter on screw pullout resistance. Acta Ortop Bras 2007;15(2):76-9.
Cho KC, Baek SH. Effects of predrilling depth and implant shape on the mechanical properties of orthodontic mini-implants during the insertion procedure. Angle Orthod 2011;82(4):618-24.