Comparison of Setup Errors in Nasopharyngeal Cancer Radiotherapy With and Without a Shoulder Retractor Using Three-Dimensional CBCT Data
Keywords:
patient positioning, setup error, nasopharyngeal cancer, shoulder retractor, cone-beam computed tomography (CBCT)Abstract
Background: Accurate patient positioning is essential in radiotherapy for nasopharyngeal cancer because the target volume is located adjacent to critical organs. Although cone-beam computed tomography (CBCT) improves pretreatment position verification, movement of the neck and shoulder regions remains a potential source of setup error. A shoulder retractor is commonly used to reduce shoulder elevation during treatment; however, evidence regarding its impact on setup accuracy remains limited.
Objective: To compare setup errors between nasopharyngeal cancer patients treated with and without a shoulder retractor using three-dimensional CBCT-based position verification.
Materials and Methods: This retrospective analytical study included 36 patients with nasopharyngeal cancer treated using a TrueBeam linear accelerator (LINAC) of Varian. Patients were divided into two groups: those treated with a shoulder retractor (SR; n = 18) and those treated without a shoulder retractor (No-SR; n = 18). CBCT images from at least four treatment fractions per patient were analyzed. The mean translational setup errors in the lateral, longitudinal, and vertical directions were calculated as the per-patient mean setup error. Total vector error (TVE) was also calculated. Differences between the two groups were analyzed using an independent two-sample t-test. Statistical significance was defined as p < 0.05.
Results: The mean setup errors in the SR group were 0.33 ± 0.87 mm, 0.29 ± 1.02 mm, and -0.54 ± 1.10 mm in the lateral, longitudinal, and vertical directions, respectively. The corresponding values in the No-SR group were 0.04 ± 0.69 mm, -0.03 ± 1.10 mm, and -0.15 ± 1.04 mm, respectively. The mean TVE was 1.63 ± 0.84 mm in the SR group and 1.48 ± 0.68 mm in the No-SR group. No statistically significant differences were observed between the two groups for any of the evaluated parameters (p > 0.05).
Conclusion: No statistically significant differences in setup errors were observed between patients treated with and without a shoulder retractor based on CBCT evaluation. Within the clinical setting of this study, the use of a shoulder retractor showed no clear advantage in reducing setup errors during radiotherapy for nasopharyngeal cancer.
References
1. Hirotaki K, Tomizawa K, Moriya S, Ito M, Sakae T. Impact of anatomical position errors on dose distribution in head and neck radiotherapy and robust image registration against anatomical changes. Anticancer Res. 2023; 43:1827-34.
2. Lenko H, Peterlin P. The impact of reference isocentre position on set-up errors in head-and-neck image-guided radiotherapy. J Radiother Pract. 2018; 17:104-13.
3. Park ET, Park SK. Setup uncertainties for inter-fractional head and neck cancer in radiotherapy. Oncotarget. 2016; 7:46662-7.
4. Contesini M, Guberti M, Saccani R, Braglia L, Iotti C, Botti A, et al. Setup errors in patients with head-neck cancer (HNC), treated using the intensity modulated radiation therapy (IMRT) technique: how it influences the customised immobilisation systems, patient's pain and anxiety. Radiat Oncol. 2017; 12:72.
5. Androjna S, Marcius VZ, Peterlin P, Strojan P. Assessment of set-up errors in the radiotherapy of patients with head and neck cancer: standard vs. individual head support. Radiol Oncol. 2020; 54:364-70.
6. Newbold KL, Bhide S, Convery H, Harrington KJ, Nutting CM. Prospective intra-patient evaluation of a shoulder retraction device for radiotherapy in head and neck cancer. Med Dosim. 2012; 37:293-5.
7. Neubauer E, Dong L, Followill DS, Garden AS, Court LE, White RA, et al. Assessment of shoulder position variation and its impact on IMRT and VMAT doses for head and neck cancer. Radiat Oncol. 2012; 7:19.
8. van Lin EN, van der Vight L, Huizenga H, Kaanders JH, Visser AG. Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customised head and neck support. Radiother Oncol. 2003; 68:137-48.
9. Bell K, Licht N, Rube C, Dzierma Y. Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment. Radiat Oncol. 2018; 13:190.
10. Den RB, Doemer A, Kubicek G, Bednarz G, Galvin JM, Keane WM, et al. Daily image guidance with cone-beam computed tomography for head-and-neck cancer intensity-modulated radiotherapy: a prospective study. Int J Radiat Oncol Biol Phys. 2010; 76:1353-9.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Thai Association of Radiation Oncology

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
บทความที่ได้รับการตีพิมพ์เป็นลิขสิทธิ์ของวารสารมะเร็งวิวัฒน์ ข้อความที่ปรากฏในบทความแต่ละเรื่องในวารสารวิชาการเล่มนี้เป็นความคิดเห็นส่วนตัวของผู้เขียนแต่ละท่านไม่เกี่ยวข้องกับ และบุคคลากรท่านอื่น ๆ ใน สมาคมฯ แต่อย่างใด ความรับผิดชอบองค์ประกอบทั้งหมดของบทความแต่ละเรื่องเป็นของผู้เขียนแต่ละท่าน หากมีความผิดพลาดใดๆ ผู้เขียนแต่ละท่านจะรับผิดชอบบทความของตนเองแต่ผู้เดียว