Impact of iterative Metal Artifact Reduction (iMAR) on External Beam Radiotherapy: A Phantom Sttudy

Authors

  • Ploy Dangsamak 1Division of Radiation Oncology, Department of Radiology, Faculty of Medicine Siriraj Hospital, Mahidol University
  • Sansanee Kongkum Division of Radiation Oncology, Department of Radiology, Faculty of Medicine Siriraj Hospital, Mahidol University
  • Pitchayut Nakkrasae Division of Radiation Oncology, Department of Radiology, Faculty of Medicine Siriraj Hospital, Mahidol University

Keywords:

Iterative Metallic artifact reduction (iMAR), Hounsfield unit correction, Dose calculation accuracy

Abstract

Background: Metallic implants introduce significant artifacts in computed tomography (CT) images, leading to Hounsfield Unit (HU) inaccuracies and subsequent errors in external beam radiotherapy dose calculations.

Objective: This study evaluated the impact of iterative metal artifact reduction (iMAR) on CT image quality, electron density representation, and dosimetric accuracy using a phantom-based model.

Materials & Methods: Initial CT simulator performance was validated against AAPM Task Group 66 standards using a Catphan® 600 phantom. To assess HU recovery, carbon steel rods were inserted into a CIRS Gammex phantom in various configurations to simulate clinical scenarios (e.g., spinal lesions, prosthetic hips). Images were reconstructed with and without the iMAR algorithm across five smoothness levels. Dosimetric impact was evaluated using a CIRS pelvic phantom and the Eclipse treatment planning system. Dose distributions for 6 MV and 10 MV photon beams were calculated using the Grid-Based Boltzmann Solver (GBBS) algorithm. Accuracy was quantified using point dose comparisons and gamma index analysis.

Results: CT quality assurance confirmed all parameters were within clinical tolerances. In the presence of metal, uncorrected images exhibited severe streaking and photon starvation. However, iMAR significantly restored HU fidelity and anatomical visualization, particularly in complex multi-metal geometries. Dosimetrically, iMAR-corrected datasets yielded point doses closer to the reference data, reducing the percentage dose difference (from 0.66% to 0.44% at 6 MV and 1.82% to 0.64% at 10 MV). While both corrected and uncorrected plans passed the 3%/3 mm gamma criteria (>99%), the stricter 2%/2 mm criteria revealed a substantial decline in uncorrected passing rates (~87–89%). The application of iMAR restored these passing rates to >97% and significantly reduced standard deviation, indicating improved procedural consistency.

Conclusion: The iMAR algorithm effectively mitigates metal-induced artifacts and restores CT number accuracy. Its implementation is critical for high-precision EBRT planning, as it ensures dose calculation stability and superior gamma passing rates under stringent clinical criteria.

References

1. Verburg JM, Seco J. CT metal artifact reduction methods: A quantitative comparison of commercial and non-commercial techniques. Med Phys. 2012;39: 4367–79.

2. Reft C, Alecu R, Das IJ, Gerbi JB, Keall P, Lief E, et al. Dosimetric considerations for patients with hip prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63. Med Phys. 2003;30:1162–82.

3. Boas FE, Fleischmann D. CT artifacts: Causes and reduction techniques. Imaging Med. 2012;4:229–40.

4. Siemens Healthineers. Iterative Metal Artifact Reduction (iMAR): Technical principles and clinical applications. Siemens White Paper; 2015.

5. Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieß M. Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys. 2010;37:5482–93.

6. Axente M, Paidi A, Von Eyben R, Hancock SL, Skinner L, Watkins WT, et al. Clinical evaluation of metal artifact reduction for CT-based radiotherapy planning. Phys Med Biol. 2015;60:4491–511.

7. Zhang Y, Huang J, Chen W, Chen J, Yan G, Lu J. Dosimetric evaluation of metal artifact reduction algorithms in radiotherapy planning using phantom studies. Phys Med. 2016;32: 235–43.

8. Mutic S, Palta JR, Butker EK, Das IJ, Huq MS, Loo LN, et al. Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66. Med Phys. 2003;30:2762-92.

9. Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology. 1987;164:576-7.

10. Maerz M, Koenig S, Schmitt D, Ackermann B, Ellerbrock M, Greilich S. Metal artifact reduction in CT imaging for radiotherapy planning: Dosimetric effects and clinical relevance. Phys Med Biol. 2016;61:N298-312.

11. Schmitt D, Maerz M, Koenig S, Ackermann B, Ellerbrock M, Greilich S. Impact of iterative metal artifact reduction on dose calculation accuracy in radiotherapy. Phys Med. 2021;82:1-8.

Downloads

Published

2026-10-28

How to Cite

1.
Dangsamak P, Kongkum S, Nakkrasae P. Impact of iterative Metal Artifact Reduction (iMAR) on External Beam Radiotherapy: A Phantom Sttudy. J Thai Assn of Radiat Oncol [Internet]. 2026 Oct. 28 [cited 2026 Nov. 7];32(2):R1-R15. Available from: https://he01.tci-thaijo.org/index.php/jtaro/article/view/286914