Surface dose measurements in clinical 6 MeV-X-photon beams : Comparison of Measurements and MC calculations
Abstract
For radiotherapy of superfi cial lesions, but also in intensity modulated radiotherapy (IMRT), 6 MeV-X-photon beams are preferred, with relative surface doses, dependent on fi eld size and SSD, of 8 % to 28 %. Depending on absorption material within the beam this values can vary considerable. The infl uence of a block tray, different metals and a new type of transmission detector on surface doses as well as on other beam parameters have been investigated. The surface doses as well as some other fi eld parameters have been checked by means of a plane parallel ionization chamber (Markus chamber, type 23 343, PTW-Freiburg, Germany). Because of specifi c design features the readings have to be corrected, and the results have been verifi ed by Monte Carlo (MC) simulations and calculations: Measurements agreed well with calculations within Δ ≈ 3 %. So it could be shown for the used Markus chamber, type 23 343, that the applied correction function is suitable for open fi elds as well as for blocked fi elds at different clinical used SSDs.
References
S. Kim, C. R. Liu, et al : Photon beam skin dose analysis for different clinical setups. Med. Phys. 25, 860-866 (1998)
D. E. Mellenberg: Determination of build-up region over-response corrections for a Markus-type chamber. Med. Phys. 17, 1041-1044 (1990)
W. Abdel-Rahman, J. P. Seuntjens, et al : Validation of Monte Carlo calculated surface doses for megavoltage photon beams. Med. Phys. 32, 286-98 (2005)
CRC Handbook of Chemistry and Physics (Ed. D. R. Lide) , CRC Press, Boca Raton 2000
IBA Information-brochure for "COMPASS", Schwarzenbruck, Germany, 2007
S. Venkataraman, K. Malkoske, M. Jensen, et al : The infl uence of a novel transmission detector on 6 MV x-ray beam characteristics. Phys. Med. Biol. 54, 3171-83 (2009)
PTW-Freiburg, Detector-catalog, Freiburg, Germany, 2007
DIN 6800-2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. Beuth Verlag Berlin, 2008
H. Reich (Hrg.) : Dosimetrie ionisierender Strahlung. B. G. Teubner, Stuttgart 1990
J. M. Jensen, G. Asuni, I. Lütjens, B. McCurdy : Oberfl ächendosisbestimmung in klinischen 6 MeV X-Photonenfeldern. mt-medizintechnik 129, 144-147 (2009)
R. G. Jaeger, W. Hübner (Eds.): Dosimetrie und Strahlenschutz. G. Thieme Verlag, Stuttgart 1974
Downloads
Published
How to Cite
Issue
Section
License
บทความที่ได้รับการตีพิมพ์เป็นลิขสิทธิ์ของวารสารมะเร็งวิวัฒน์
ข้อความที่ปรากฏในบทความแต่ละเรื่องในวารสารวิชาการเล่มนี้เป็นความคิดเห็นส่วนตัวของผู้เขียนแต่ละท่านไม่เกี่ยวข้องกับ และบุคคลากรท่านอื่น ๆ ใน สมาคมฯ แต่อย่างใด ความรับผิดชอบองค์ประกอบทั้งหมดของบทความแต่ละเรื่องเป็นของผู้เขียนแต่ละท่าน หากมีความผิดพลาดใดๆ ผู้เขียนแต่ละท่านจะรับผิดชอบบทความของตนเองแต่ผู้เดียว