Abstract

Background: It is predictable that the laboratories with many years of accreditation should get lower amount of nonconformities than those of the new ones. The objectives of the study are to find the differences in characteristics and nonconformities among ISO/IEC 17025 accredited laboratories undergoing on-site assessment in 2015 and verify the above prediction.


Materials and methods: A cross-sectional study of the characteristics and the nonconformities from 2015 assessment of 134 laboratories were compared between the laboratories with 6 or more years or at least 3 cycles of accreditation and the ones with shorter accreditation year. Statistical analyses were done by using Stata-intercooled version 6.


Results: There were 1,459 total nonconformities, 572 (39%) were those against management requirements and 887 (61%) were those from technical ones. Longer and new accredited laboratories shared similar basic characteristics. Longer accredited laboratories got lower total nonconformities (mean±SD; 9.1±4.8 vs 12.7±9.0: t-test=2.89, p=0.003), and lower technical nonconformities (mean±SD; 5.0±3.1 vs 8.3±6.9: t-test=3.57, p<0.001). The proportion of longer accredited laboratories was lower on nonconformities against clause 5.6 Measurement traceability (14/87 vs 28/67, Chi square=5.060, p=0.024) and the ones against clause 5.9 Assuring quality of test results (27/67 vs 44/67, Chi-square=6.450, p=0.011) than those of the new accredited laboratories.


Conclusion: This study demonstrated that many year accreditations may enable the accredited laboratories to comply with the requirements more effectively. Lower proportions of the longer accredited laboratories had lower certain nonconformities on clause 5.6 Measurement traceability and 5.9 Assuring quality of test results.

Article Details

How to Cite
Vongsheree, S., Aromsuk, S., Chainarongkuekul, P., & Kejonnit, A. (2018). Lower frequency of certain nonconformities against ISO/IEC 17025 after many year accreditation. Journal of Associated Medical Sciences, 51(1), 19–25. Retrieved from https://he01.tci-thaijo.org/index.php/bulletinAMS/article/view/91330
Section
Research Articles
Author Biography

Suthon Vongsheree, Department of Medical Sciences, Ministry of Public Health, Nonthaburi Province, Thailand

Senior Medical Scientist

References

[1] International Laboratory Accreditation Cooperation. (2016). ILAC MRA and signatories. Retrieved from: https://ilac.org/signatory-search/?q=b/.

[2] International Organization for Standardization. ISO/IEC 17025:2005.General requirements for the competence of testing and calibration laboratories. Geneva, Switzerland 2005.

[3]Aromsuk S, Sinthusarn C, Tanankat A and Jarunuch S. Trend analysis of condition on accredited assessment for public health laboratory of government and public laboratories service in accordance with ISO/IEC 17025:2005 from the fiscal year 2008 to 2011.Rajabhat Chiang Mai Research Journal.2013: 14(2).English abstract.

[4] Bella Ho and Eric Ho.The most common nonconformities encountered during the assessment of medical laboratories in Hong Kong using ISO 15189 as accreditation criteria. Biochem Med (Zagreb) 2012: 22(2): 247-57.

[5]Aromsuk S, Chainarongkuekul P, Khejonnit A and Vongsheree S. Factors affecting the duration of the correction time for public health laboratory and veterinary laboratory in accordance with ISO/IEC 17025:2005.Journal of Applied Animal Science. 2016:9(1): 9-24. English abstract.

[6] Buthmann A. Dealing with Non-normal Data: Strategies and Tools.Available at https://www.isixsigma.com/tools-templates/normality/dealing-non-normal-data-strategies-and-tools/
[7] Muenphon S, Wiroonboonpat P, Kumsai S, et al. Factors affecting the numbers of nonconformities from the assessment of safety in medical laboratory according to ISO 15190:2003.Bull Chaing Mai Assoc Med Sci 2016: 49(2): 255-62. English abstract.

[8] International Organization for Standardization.ISO/IEC 17011:2004.Conformity assessment-General requirements for accreditation bodies accrediting conformity assessment bodies. Geneva, Switzerland 2004.