A Comparison of The Temperature Measure by Nasal Temperature Probe, Infrared Forehead Skin Thermometer and Infrared Tympanic Thermometer in General Anesthesia Patients
Keywords:
a nasal tube body thermometer, an infrared forehead skin thermometer, an infrared ear thermometer, surgical patientsAbstract
Background: During a previous COVID-19 virus outbreak, inserting a nasal probe to measure temperature in surgical patients undergoing general anesthesia posed a risk of infection spread. Non-contact body temperature measurement is therefore an alternative method for monitoring the body temperature of patients during surgery.
Objective: to compare body temperature measure by the nasal core temperature method, infrared forehead skin thermometer , and infrared tympanic thermometer
Materials and Methods: This quasi-experimental research involved 216 surgical patients under general anesthesia. The tools used in the research were: 1) The experimental tool was a nasal tube body thermometer, a forehead skin thermometer, and an infrared tympanic thermometer. 2) a recorder of data. By simultaneously measuring the temperature using all three methods and taking readings
Results: A nasal tube body thermometer, a forehead skin thermometer, and an infrared tympanic thermometer in surgical patients receiving general anesthesia at 1 and 30 minutes were significantly different when considering the pairwise temperature difference between the nasal and tympanic core temperature measurements. The mean temperature was found to be not significantly different (p =.787). The result of temperature difference between nasal and skin, tympanic and skin were significantly different ( p < 0.001 )
Conclusions: Tympanic temperature measurements in surgical patients undergoing general anesthesia are similar to nasopharyngeal temperature measurements commonly used in surgical patients.
Keywords: a nasal tube body thermometer, an infrared forehead skin thermometer, an infrared ear thermometer, surgical patients
References
Hymczak H, Gołąb A, Mendrala K, Plicner D, Darocha T, Podsiadło P, et al. Core Temperature Measurement-Principles of Correct Measurement, Problems, and Complications. Int J Environ Res Public Health 2021;18(20): 1-8.
Poveda VdB, Nascimento AdS. Intraoperative body temperature control: esophageal thermometer versus infrared tympanic thermometer. Rev Esc Enferm USP 2016;50:946-52.
Yeoh WK, Lee JKW, Lim HY, Gan CW, Liang W, Tan KK. Re-visiting the tympanic membrane vicinity as core body temperature measurement site. Plos One 2017;12(4):e0174120.
ราชวิทยาลัยวิสัญญีแพทย์แห่งประเทศไทย [อินเตอร์เน็ต]. ประกาศราชวิทยาลัยวิสัญญีแพทย์แห่งประเทศไทย ปี พ.ศ.๒๕๖๒ เรื่อง แนวทางพัฒนาการให้ข้อมูลเกี่ยวกับการระงับความรู้สึก [สืบค้นเมื่อ 1 สิงหาคม 2565] เข้าถึงจาก http://www.anesthai.org/public/rcat/Documents /document/ 1571131341 %B5%E0%B9%88%E0%B8%A2%E0%B8%A7%E0%B8%81% E0%B8%B1%E0%B8%9A %E0%B8%81%E0%B8%B2%E0%B8%A3.pdf
มนสิชา สมจิตร, นรินทร์ พลายละหาร, อัครวัฒน์ สินเกื้อกูลกิจ, วิลาวัลย์ สมดี, วิริยา ถิ่นชีลอง, พุ่มพวง สาระพาณิชย์. การศึกษาเปรียบเทียบอุณหภูมิแกนของผู้ป่วยที่ใช้ Forced-Air Warming ด้วยวิธี Modified Lower-Body Cover กับ Commercial Lower-Body Cover ในระหว่างการผ่าตัดช่องท้องขนาดใหญ่. ศรีนครินทร์เวชสาร 2564;36(4):401-8.
Yang F, Wang J, Cui J, Zhuan J, Hu X, Chen S. An Overview of the Implications for Perianesthesia Nurses in terms of Intraoperative Changes in Temperature and Factors Associated with Unintentional Postoperative Hypothermia. J Healthc Eng 2022;6955870. Published online 2022 Apr 11 doi: 10.1155/2022/6955870.
Mogensen CB, Wittenhoff L, Fruerhøj G, Hansen S. Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes. BMC Pediatr 2018;18(1):15. doi: 10.1186/s12887-018-0994-1.
Ramirez-GarciaLuna JL, Bartlett R, Arriaga-Caballero JE, Fraser RDJ, Saiko G. Infrared Thermography in Wound Care, Surgery, and Sports Medicine: A Review. Front Physiol 2022;13:838528. doi: 10.3389/fphys.2022.838528.
Wang L, Yin H, Di Y, Liu Y, Liu J. Human local and total heat losses in different temperature. Physiol Behav 2016;157:270-6.
Hine K, Hosono S, Kawabata K, Miyabayashi H, Kanno K, Shimizu M, Takahashi S. Nasopharynx is well-suited for core temperature measurement during hypothermia therapy. Pediatr Int 2016;59(1):29-33.
Wan L, Shen PY, Zhang SX, Wang LZ. Agreement of infrared ear temperature with nasopharyngeal temperature and diagnostic performance on hypothermia in general anesthetized patients. J Chin Med Assoc 2022;85(11):1093-7.
Songlin Y, Hailiang Z, Wei W, Zhe W, editors. A novel infrared forehead thermometer with ambient temperature compensation for body temperature measurement in low-temperature environment. 5th Optics Young Scientist Summit 2022;12448. https://doi.org/10.1117/12.2626262
Bräuer A, Fazliu A, Brandes IF, Vollnhals F, Grote R, Menzel M. Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature. J Clin Monit Comput 2023;37(1):29-36.
Machin G, Brettle D, Fleming S, Nutbrown R, Simpson R, Stevens R, et al. Is current body temperature measurement practice fit-for-purpose? J Med Eng Technol 2021;45(2):136-44.
Fletcher T, Whittam A, Simpson R, Machin G. Comparison of non-contact infrared skin thermometers. J Med Eng Technol 2018;42(2):65-71.
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
เนื้อหาและข้อมูลในบทความที่ลงตีพิมพ์ในมหาราชนครศรีธรรมราชเวชสาร ถือเป็นข้อคิดเห็นและความรับผิดชอบของผู้เขียนบทความโดยตรงซึ่งกองบรรณาธิการวารสาร ไม่จำเป็นต้องเห็นด้วย หรือร่วมรับผิดชอบใด ๆ
บทความ ข้อมูล เนื้อหา รูปภาพ ฯลฯ ที่ได้รับการตีพิมพ์ในมหาราชนครศรีธรรมราชเวชสาร ถือเป็นลิขสิทธิ์ของมหาราชนครศรีธรรมราชเวชสาร หากบุคคลหรือหน่วยงานใดต้องการนำทั้งหมดหรือส่วนหนึ่งส่วนใดไปเผยแพร่ต่อหรือเพื่อกระทำการใด จะต้องได้รับอนุญาตเป็นลายลักษณ์อักษรจากวารสารมหาราชนครศรีธรรมราชเวชสาร ก่อนเท่านั้น