A Comparison Between Breath hold and Free Breathing MRI Techniques of Myocardial T2* Measurement

Main Article Content

Uten Yarach
Suwit Saekho
Patai Buttakote
Siriphan Luxsakhum
Arintaya Phrommintikul
Nipon Chattipakorn

Abstract

Objectives : To compare the effectiveness and reproducibility of myocardial T2* measurement between breath-hold and free-breathing techniques using GRE black blood multi-echoes pulse sequence.


Methods : The techniques were validated by a known iron concentration phantom prior to study in 15 healthy volunteers. The images of single short-axis view of the mid left ventricle were acquired at 10 echo times using both Breath-hold and Free-breathing techniques. The images from both techniques were drawn the Regions of interest (ROI) at the left ventricular septum for T2* analysis. The parameters to be analyzed include Signal to Noise Ratio (SNR), reproducibility and percent coefficient of Variation (%CV). The fitting models used to evaluate for T2* values were the mono-exponential and offset models done by MATLAB 7.01.


Results : The correlation between known iron concentration phantom and R2*(1000/T2*) using Pearson‘s test showed strongly linear correlation with a correlation coefficient over 0.96 (p<0.001). In human study, the images obtaining from the Free-breathing technique provided higher SNR than that of the Breath-hold approximately 40% in all TE images as well as the mean T2* values acquiring from different fitting models were no significant different (36.59±3.18 ms., and 36.31±2.72 ms., ( p = 0.799)), while the mean T2* values of Breath-hold technique obtaining from different fitting models were significant different (35.69±2.90 ms., and 30.67±7.21, (p = 0.036)). For the coefficients of variation (%CV), the T2* from the Free-breathing technique was lower than that of the Breath-hold technique, 6.34% and 8.41%, respectively.


Conclusion : The Free-breathing T2* technique provided higher resolution and SNR compared to that of the Breath-hold T2* technique. In addition, the Free-breathing technique showed higher reproducibility when evaluated at different time frames of study, different fitting models, and different observers.

Article Details

How to Cite
Yarach, U., Saekho, S., Buttakote, P., Luxsakhum, S., Phrommintikul, A., & Chattipakorn, N. (2010). A Comparison Between Breath hold and Free Breathing MRI Techniques of Myocardial T2* Measurement. Journal of Associated Medical Sciences, 43(3), 192. Retrieved from https://he01.tci-thaijo.org/index.php/bulletinAMS/article/view/60076