Dose deviations induced by fractional image guidance system errors in intensity-modulated radiotherapy for brain tumor treatment

Main Article Content

Siriprapa Somboon
Wannita Malila
Surasak Tamon
Jumneanphan Rueansri
Supoj Ua-apisitwong

Abstract

Background: Intensity-modulated radiotherapy (IMRT) techniques have a steep dose distribution, leading to large dosimetric errors caused by incorrect daily patient setup. Image-guided radiotherapy (IGRT) is an essential technique to verify accurate patient setup. The accuracy of verifying a patient’s position depends on many factors including the image guidance system and image registration software.


Objectives: This study investigated the dosimetric and geometric differences between the original plan and simulated plans with setup errors associated with kilovoltage cone-beam computed tomography (kV-CBCT), kV, kV planar image, and electronic portal imaging detector (EPID) for brain tumor treatment using IMRT.


Materials and methods: A PIXY Anthropomorphic Training/Teaching Phantom was used in this study. IMRT treatment plans with five co-planar fields from a 6 MV Elekta Versa HD linear accelerator were generated using the RayStation computer treatment planning system. Three image guidance systems, including kV-CBCT, kV planar imager, and EPID were used to perform image registration. To evaluate the efficiency of each image guidance system, a simulated setup error by couch was shifted 0, ±2, and ±4 mm in the lateral, longitudinal, and vertical planes. Errors in the image registration from the actual couch shift were collected. Measured errors were used to generate the treatment plans of 25 fractions using the IMRT technique by random shifts of the isocenter of each fraction while maintaining other planning parameters of the original plan. The dose deviation in planning target volume (PTV) and geometric deviation compared to original plan were recalculated and analyzed.


Results: Accuracy of image registration in all image guidance systems indicated that the registration errors were less than 1.7, 2.0, and 1.0 mm for the kV-CBCT, kV planar, and EPID, respectively. The average PTV dosimetric deviation induced by the setup error ranged from -2 to 2 mm per fraction; this study showed dosimetric deviation at D98% approximately -6%, D95% approximately -4% and D2% below -1%, while the average of isodose distribution shift of -5% inside the PTV of the isocenter region were -1.58±0.67, -1.28±0.52 and -1.30±0.80 mm in the cerebella, parasagittal, and convex areas, respectively. An isodose shift of -10% was <1 mm in all PTV locations.


Conclusion: The efficiency of image guidance resulted in small errors within ±2 mm using kV-CBCT, kV planar, and EPID. The setup error influenced daily dose distribution, while the PTV recorded underdose of about 4% in the brain IMRT technique. 

Article Details

How to Cite
Somboon, S., Malila, W., Tamon, S., Rueansri, J., & Ua-apisitwong, S. (2022). Dose deviations induced by fractional image guidance system errors in intensity-modulated radiotherapy for brain tumor treatment. Journal of Associated Medical Sciences, 55(3), 116–122. Retrieved from https://he01.tci-thaijo.org/index.php/bulletinAMS/article/view/256065
Section
Research Articles

References

Marta GN, Silva V, de Andrade Carvalho H, de Arruda FF, Hanna SA, Gadia R, et al. Intensity-modulated radiation therapy for head and neck cancer: systematic review and meta-analysis. Radiother Oncol. 2014; 110(1): 9-15.

Dawson LA, Jaffray DA. Advances in image-guided radiation therapy. J Clin Oncol. 2007; 25(8): 938-46.

Bissonnette JP, Purdie TG, Higgins JA, Li W, Bezjak A. Cone beam computed tomographic image guidance for lung cancer radiation therapy. Int J Radiat Oncol Biol Phys. 2009; 73(3): 927-34.

Snir JA, Battista JJ, Bauman G, Yartsev S. Evaluation of inter-fraction prostate motion using kilovoltage cone beam computed tomography during radiotherapy. Clin Oncol. 2011; 23(9): 625-31.

Guckenberger M. Image-guided radiotherapy based on kilovoltage cone-beam computed tomography – A review of technology and clinical outcome. Eur Oncol Haematol. 2011; 7(2): 121-4.

Pisani L, Lockman D, Jaffray D, Yan D, Martinez A, Wong J. Setup error in radiotherapy: on-line correction using electronic kilovoltage and megavoltage radiographs. Int J Radiat Oncol Biol Phys. 2000; 47(3): 825-39.

Somboon S, Malila W, Tamon S, Nueangwong W, Yeenang N, Rueansri J. Evaluation of optimal kilovoltage-cone beam technique on image quality, registration accuracy, time of imaging and relative dose for head radiotherapy: A phantom study. J Assoc Med Sci. 2022; 55(2): 10-5.

Devereux B, Frantzis J, Sisson T, Jones M, Martin J, Middleton M. A comparison of kV and imaging in head and neck imaged radiotherapy. Radiography. 2010; 16: 8-13.

Martin L, Couto JG, Barbosa B. Use of planar kV vs. CBCT in evaluation of setup errors in oesophagus carcinoma radiotherapy. Rep Pract Oncol Radiother. 2016; 21: 57-62.

Veerasarn K, Yuthagovit S, Chailorrat RN. Prevalence of Brain tumor in Thailand from 2005 to 2014: data from the National Health Security Office. J Med Assoc Thai. 2016; 99 Suppl3: 62-73.

Li Z, Shen D, Zhang J, Zhang J, Yang F, Kong D et al. Relationship between WBRT total dose, intracranial tumor control, and overall survival in NSCLC patients with brain metastases- a single-center retrospective analysis. BMC cancer. 2019; 19(1): 1104.

Li J, Bentzen SM, Renschler M, Mehta MP. Regression after whole-brain radiation therapy for brain metastases correlates with survival and improved neurocognitive function. J Clin Oncol. 2007; 25(10): 1260-6.

Ding XG, Alaei P, Curran B, Flynn R, Gossman M, Mackie TR et al. Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180. Med Phys. 2018; 45(5): e84-e99.

Brock KK, Mutic S, McNutt RT, Li H, Kessler LM. Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No.132. Med Phys. 2017; 44(7): e43-76.

Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, Flentje M. Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int J Radiat Oncol Biol Phys. 2006; 65(3): 934-42.

Utena Y, Takatsu J, Sugimoto S, Sasai K. Trajectory log analysis and cone-beam CT-based daily dose calculation to investigate the dosimetric accuracy of intensity-modulated radiotherapy for gynecologic cancer. J Appl Clin Med Phys. 2021; 22(2): 108-17.

Siebers JV, Keall PJ, Wu Q, Williamson JF, Schmidt-Ullrich RK. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. Int J Radiat Oncol Biol Phys. 2005; 2: 422-33.

Group RTO. A phase II study of intensity modulated radiation therapy (IMRT) +/−chemotherapy for nasopharyngeal cancer, RTOG 0225[2011-03-05]. http://208.251.169.72/members/protocols/0225/pdf_fle.html.

Cubillos Mesias M, Boda-Heggemann J, Thoelking J, Lohr F, Wenz F, Wertz H. Quantification and Assessment of Interfraction Setup Errors Based on Cone Beam CT and Determination of Safety Margins for Radiotherapy. PLoS One. 2016; 11(3): e0150326.

International Commission on Radiation Units and Measurements. Prescribing, Recording, and Reporting Photon Beam Therapy ICRU Report 50, ICRU Report 62 (Supplement to ICRU Report 50) 1993. Bethesda:ICRU,1999.

Bell K, Licht N, Rube C, Dzierma Y. Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment. Radiat Oncol. 2018; 13(1): 190.