Reliability of Portable Blood Glucose Meters for Using in Small Animal Hospital

Authors

  • Siram Suvarnavibhaja
  • Poramin Yuennan
  • Ratanaporn Tangwangwiwat
  • Aomusa Kuaha
  • Sirintorn Yibchokeanan

Keywords:

dog, glucose concentration, portable blood glucose meter

Abstract

The objective of this study was to compare blood glucose concentrations obtained from two portable blood glucose meters (PBGM), Sure-step and Accucheck, with reference method, glucose oxidase test. Blood samples were collected from dogs consulting in Small Animal Hospital, Chulalongkorn University. A hundred and thirty four blood samples were collected in the study. Blood glucose concentrations were measured by using Sure-step™, Accucheck™ and glucose oxidase method respectively in all samples. The intra-class correlation was determined to evaluate the correlation between each PBGM and glucose oxidase test. The range of the blood glucose concentrations was 31.47 to 347.49 mg/dl. Overall intra-class correlation showed significant (p<0.05) in comparison of two PBGMs with the standard method (r=0.847 and 0.839 for Sure-step™ and Accucheck™ respectively).  Range of 15% deviation from reference method-value was accepted to access PBGMs clinically.   The greater percentage of sample in particular range was observed in Sure-step™ (42.86%) rather than Accucheck™ (22.56%). In conclusion, although commercially available PBGM provided blood glucose concentration reasonably close to those obtained with reference method, Sure-step™ seemed to be more accurate than Accucheck™. However the utilization of these devices in terms of clinical decision could lead to erroneous in some cases.

Downloads

How to Cite

Suvarnavibhaja, S., Yuennan, P., Tangwangwiwat, R., Kuaha, A., & Yibchokeanan, S. (2014). Reliability of Portable Blood Glucose Meters for Using in Small Animal Hospital. The Thai Journal of Veterinary Medicine, 44(2), 195–200. Retrieved from https://he01.tci-thaijo.org/index.php/tjvm/article/view/18566

Issue

Section

Original Articles